STEP 05 | remove transportation lock

 

If the device hangs upright on the wall, you can remove the transportation lock in the reagent cartridge receptacle.

The transport lock in the reagent cartridge receptacle should prevent you from losing the magnetic stirrer inside the device.

Please remove the transport lock not before the device hangs upright on the wall.

 

Colormetry CMU 324 transport lock 1

 

NOTE 2The magnetic stirrer must be positioned on bottom of the cylindric part of the reagent cartridge receptacle.
If the Colormetry is operated without the magnetic stirrer, monitoring cannot be performed properly and an error message
"F281: incomplete reagent injection" will be displayed. 
In case of lost, a replacement stirrer is provided on the backside of the unit`s cover plate.

  
 
  Colormetry CMU 324 HE magnetic stirrer 004Colormetry CMU 324 HE magnetic stirrer 003

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.